Transfer Function Analyser TFA. 46

Servomex Controls Ltd.; Crowborough, Sussex

  • Anno
  • 1959 ?
  • Categoria
  • Strumento da laboratorio
  • Radiomuseum.org ID
  • 372503

 Specifiche tecniche

  • Gamme d'onda
  • - senza
  • Tensioni di funzionamento
  • Alimentazione a corrente alternata (CA)
  • Altoparlante
  • - - Nessuna uscita audio.
  • Materiali
  • Mobile di metallo
  • Radiomuseum.org
  • Modello: Transfer Function Analyser TFA. 46 - Servomex Controls Ltd.;
  • Forma
  • Rack
  • Annotazioni
  • Flagship ultra-low-frequency transfer function analyser for measuring the gain and phase (in-phase and quadrature) response of automatic control systems, servomechanisms, and industrial process plants.
    Housed in a four-tier 19-inch enclosed rack cabinet featuring twin rectangular panel meters at the top, a central oscillator/resolver control unit, and lower power-supply and electro-mechanical drive bays.
    Listed in the 18 September 1961 Servomex Price List at £2,300 U.K. delivered (£2,340 Export), making it the highest-priced instrument in the firm's catalogue.

    Operating Principle & Outputs:
    Designed to deliver steady readings down to wave periods of 10,000 seconds (2¾ hours / 0.0001 Hz) with high rejection of harmonics and system noise by resolving the returning system signal into in-phase (sinωt) and quadrature (cosωt) components displayed on the twin upper panel meters. To test electrical, carrier-servo, and mechanical control loops directly, the instrument provides three simultaneous stimulus outputs:

    • Normal sinewave output (DC-coupled ultra-low-frequency electrical sinewave)
    • Modulated carrier output (for AC synchro/magslip servo systems)
    • Mechanical output (rotary shaft drive coupled to the low-frequency generator mechanism)

    Historical Context & Modern Equivalent:
    In modern test instrumentation, the electro-mechanical transfer function analyser has evolved into the Frequency Response Analyser (FRA)—also known as a Bode Plot Analyser or Gain-Phase Analyser (such as the later British Solartron 1250/1255/1260 series). While modern FRAs and oscilloscope Bode-plot functions replace thermionic valves, magslip resolvers, and analog meters with digital signal processing (DSP/FPGA Discrete Fourier Transform correlation) and automated Gain/Phase plotting, they rely on the exact same underlying sine/cosine correlation principle to reject harmonics and noise in closed-loop control testing.

  • Prezzo nel primo anno
  • 2,300.00 £
  • Bibliografia
  • -- Original prospect or advert (Engineering, Vol. 187 (Jan–Jun 1959), p. 70.)
  • Letteratura / Schemi (1)
  • -- Original prospect or advert (Electronic Technology, November 1961, p. 7.)
  • Autore
  • Modello inviato da Gary Cowans. Utilizzare "Proponi modifica" per inviare ulteriori dati.

 Collezioni | Musei | Letteratura