EMI Test Receiver ESIB 7

Rohde & Schwarz, PTE; München

  • Year
  • 2004 ??
  • Category
  • Service- or Lab Equipment
  • Radiomuseum.org ID
  • 292043
    • alternative name: Messgerätebau Memmingen || Physikalisch-Technisches Entwicklungslabor Dr. Rohde & Dr. Schwarz

 Technical Specifications

  • Number of Transistors
  • Semiconductors present.
  • Semiconductors
  • Main principle
  • DSP, Digital Signal Processor
  • Wave bands
  • Wave Bands given in the notes.
  • Power type and voltage
  • Alternating Current supply (AC) / 100-120, 50-400Hz; 200-240, 50-60Hz Volt
  • Loudspeaker
  • - - No sound reproduction output.
  • Material
  • Metal case
  • from Radiomuseum.org
  • Model: EMI Test Receiver ESIB 7 - Rohde & Schwarz, PTE; München
  • Shape
  • Boatanchor (heavy military or commercial set >20 kg).
  • Dimensions (WHD)
  • 435 x 236 x 570 mm / 17.1 x 9.3 x 22.4 inch
  • Notes
  • Specifications in brief

    Frequency range
    Input 1: 20 Hz to 7 GHz
    Input 2: 20 Hz to 1 GHz
    Preselection in receiver mode (fixed) and analyzer mode (selectable)
    Three fixed-tuned and six or seven tracking filters (models .26 and .40)
    Preamplifier with 20 dB gain switch-selectable in conjunction with preselector

    Resolution bandwidths
    200 Hz, 9 kHz, 120 kHz in accordance with CISPR 16-1-1,
    10 Hz to 10 MHz, in decadic steps (6 dB bandwidths, receiver and analyzer mode)
    1 Hz to 10 MHz, adjustable in steps of 1/2/3/5 (3 dB bandwidths, analyzer mode)
    Parallel detectors (max. 4) in receiver mode
    Peak, average, CISPR-average, quasi-peak and RMS

    Automatic scan
    Four storable traces with up to 80 000 measured values each
    (250 000 values with one trace)

    Internal controller (Windows NT)

  • Net weight (2.2 lb = 1 kg)
  • 25.1 kg / 55 lb 4.6 oz (55.286 lb)
  • Mentioned in
  • - - Manufacturers Literature
  • Author
  • Model page created by Stefan Bode. See "Data change" for further contributors.