Frequency & Time Measuring TSA3436

Venner Electronics Ltd.; New Malden

  • Anno
  • 1964
  • Categoria
  • Strumento da laboratorio
  • Radiomuseum.org ID
  • 367649

 Specifiche tecniche

  • Numero di transistor
  • A semiconduttori.
  • Semiconduttori
  • Gamme d'onda
  • - senza
  • Tensioni di funzionamento
  • Rete / Batterie (ogni tipo) / DC: 12 V / 40 - 60 Hz, 100-125 / 200-250 Volt
  • Altoparlante
  • - - Nessuna uscita audio.
  • Materiali
  • Mobile di metallo
  • Radiomuseum.org
  • Modello: Frequency & Time Measuring TSA3436 - Venner Electronics Ltd.; New
  • Forma
  • Apparecchio portatile > 20 cm (senza la necessità di una rete)
  • Dimensioni (LxAxP)
  • 360 x 90 x 260 mm / 14.2 x 3.5 x 10.2 inch
  • Annotazioni
  • The Venner Electronics TSA3436 is a fully transistorised general‑purpose counter‑timer designed for laboratory, industrial, and electronic service‑departments in the early 1960s. It provides frequency measurement up to 1.2 Mc/s and time‑interval measurement from 1 µs to 100 000 s, using a 6‑digit in‑line numerical display for direct readout.

    Core Architecture and Frequency Standard
    At the heart of the instrument is an oven‑controlled crystal oscillator built around the Cathodeon MCO‑2M crystal oven, giving a stability of ±1 × 10⁻⁶. This temperature‑regulated reference defines all gating and counting operations. Frequency‑measurement accuracy is specified as ±1 count plus crystal stability, consistent with Venner’s transistorised decade‑counter design philosophy.

    Input Facilities
    Two selectable input modes accommodate a wide range of signal sources:

    • A.C. input: 70 kΩ // 25 pF, accepting 100 mV to 50 V r.m.s.
    • D.C. input: 1 kΩ // 500 pF, accepting 1–50 V peak.

    Three gate times—0.1 s, 1 s, and 10 s—allow the user to balance resolution against measurement speed. For external gating, the counter requires positive‑going pulses of 5–10 V, with rise times not exceeding 0.2 µs and pulse widths of at least 0.5 µs.

    Construction, Power, and Physical Data
    The TSA3436 is housed in a compact metal case measuring 14½ × 3⅝ × 10½ in and weighing approximately 13 lb. It operates from 100–125 V or 200–250 V, 40–60 c/s and can also be powered from a 12 V d.c. source for mobile or field use.

    See the 4 digit version TSA3334.

  • Peso netto
  • 13 lb (13 lb 0 oz) / 5.902 kg
  • Letteratura / Schemi (1)
  • - - Manufacturers Literature
  • Autore
  • Modello inviato da Stephen Woods. Utilizzare "Proponi modifica" per inviare ulteriori dati.

 Collezioni | Musei | Letteratura

Collezioni

Il modello Frequency & Time Measuring fa parte delle collezioni dei seguenti membri.