Catalogue de radios anciennes
D-79512
|
Pays:
Etats-Unis
|
|
| Identique à |
D-79512
|
| Similaires |
Differentes limites maxi :
D-79510
|
| Prédécesseurs |
205B
|
| Utilisé par |
Technology / Measurement |
| Chauffage |
Vf 4.5 Volt / If 1.6 Ampère / Direct |
| Description |
Ionization gauge with glass tube on top as inlet measurement port. Pear-shaped hard-glass (Nonex) version of D-79510.
As ionization gauge the anode is biased negative, -2 to -10 volts, while the grid must be positive, 100 to 125 volts typical, 5 to 25 mA. In operation filament voltage must be reduced for a grid current of 21 mA at 110 V DC. In this condition the pressure in mm. of mercury is about five times the plate current in amperes, P = 5 Ip.
Data and applications given in a PDF booklet attached to the D-79510 sensor.
|